PORTABLE MICROSCOPE SURFACE INSPECTION RECORD Yenra | Updated October 9, 2026 Guide: https://yenra.com/microscopes/portable.html Use one record per object/session. Save it beside original images and an overview/coverage sketch. This record documents observations; acceptance decisions require an applicable method and criteria. Object ID / date / observer: Inspection question and boundaries: Handling restrictions and approved contact/support: Overview filename / orientation reference: Coverage map or grid filename: Instrument / lens / zoom / working distance (with unit): Support / illumination direction and settings: Camera mode / saved dimensions in pixels / binning: Exposure and other acquisition settings: Calibration reference ID / known interval and unit: Calibration image filename / matching settings: Verification measurement of known reference: Repeatability observations / measurement uncertainty if established: Repeat for each region: Region ID / overview location: Wide-view file / detail file / repeat file: Observed feature and lighting dependence: Measured interval in pixels / conversion / result with unit: Interpretation and evidence still needed: Coverage completed / obscured or inaccessible area: ILLUSTRATIVE ARITHMETIC ONLY 2.00 mm / 1000 pixels = 0.002 mm/pixel. 150 pixels x 0.002 mm/pixel = 0.30 mm in a matching image. Use your own calibration. Recheck when spatial sampling or optics change; surface tilt, edge choice and repeatability affect the result. Final coverage gaps / next action / reviewer: Original-data location / processing history: Calibration context: https://www.leica-microsystems.com/science-lab/industrial/microscope-calibration-for-measurements-why-and-how-you-should-do-it/