PHOTOMICROGRAPH CAPTURE AND CALIBRATION RECORD Guide: https://yenra.com/microscopes/photography.html Keep the original capture and this record together. Use one record per configuration; repeat spatial calibration after changes to image sampling. Date / specimen ID / description: Microscope / objective / adapter / camera: Illumination method and settings: Exposure / gain or ISO / white balance: Capture pixel dimensions / binning / optical or digital zoom: Original image filename: Calibration slide identity and verified division size: Calibration image filename (same acquisition settings): Known distance (micrometers): Measured pixel distance: Scale = known distance / pixel distance (micrometers per pixel): Desired scale-bar length (micrometers): Bar pixels = bar length / scale: Worked example ONLY: 100 um / 400 px = 0.25 um/px; 50 um bar = 200 px. Use your own measured calibration, not the example numbers. Processing and export history (crop, resize, color, stack, other): Final output filename / calibration verified after export: Caption: specimen, method, objective, calibrated scale, consequential edits.