VIBRATION DATA ACQUISITION: MAKE MEASUREMENTS COMPARABLE Yenra | Updated September 11, 2026 Article: https://yenra.com/making/vibration.html Purpose: preserve enough acquisition context to reproduce a vibration measurement. Use with the actual sensor, conditioner and DAQ manuals. Machinery access follows approved site procedures. This record supplies no machine fault thresholds. QUESTION AND MACHINE STATE Asset / measurement question / speed rpm or other units / load / process state: Required frequency range Hz / amplitude range and convention / responsible analyst: Sensor model, serial, calibration date / sensitivity with units / range / axis: Mount location photograph / attachment / surface / torque reference / cable restraint: Conditioner/DAQ model and firmware / excitation current and compliance voltage: Sensor bias / coupling / gain / input range / expected settling behavior: Specified passband / anti-alias filter and transition / applicable manual section: Actual sample rate PER CHANNEL samples/s / channel synchronization / trigger: Sample count / observation duration s / clock or tachometer reference: FICTIONAL MATHEMATICAL CHECKS 10,240 samples/s; 20,480 actual samples => 2 s and unpadded bin spacing 0.5 Hz. Same rate; 40,960 samples => 4 s and unpadded spacing 0.25 Hz. Duration = N / sample rate; bin spacing = sample rate / N. Zero-padding adds displayed points, not additional observation time. Sensitivity 100 mV/g and 250 mV peak => 2.5 g peak. For a pure sine ONLY, 2.5 / sqrt(2) = about 1.77 g RMS. These examples assume valid linear unclipped acquisition; they do not select a machine bandwidth or diagnostic limit. VALIDATION RUN Channel identity and calibrated scaling checks: Bias/health readings / clipping / gaps / settling / range observations: Repeat at comparable machine state / differences and investigation: Original samples file / checksum or controlled record ID / settings file / photo: DERIVED ANALYSIS Software/version / window / averaging / frequency band / amplitude convention: Peak, peak-to-peak, RMS, spectrum or PSD (label exactly): Integration to velocity/displacement, if used: filters, limits and units: Result / uncertainty / analyst / date / interpretation scope: Recheck after sensor, mount, conditioning, sample plan or machine-state changes. PRIMARY REFERENCES NI's accelerometer measurement guide https://www.ni.com/en/shop/data-acquisition/sensor-fundamentals/measuring-vibration-with-accelerometers.html PCB Piezotronics' mounting guidance https://www.pcb.com/resources/technical-information/mounting anti-aliasing explanation https://www.ni.com/en/shop/data-acquisition/measurement-fundamentals/analog-fundamentals/anti-aliasing-filters-and-their-usage-explained.html LabVIEW Analysis Concepts manual https://download.ni.com/support/manuals/370192c.pdf IEPE acquisition startup saturation https://knowledge.ni.com/KnowledgeArticleDetails?id=kA03q000000YRUaCAO